Please use this identifier to cite or link to this item: http://hdl.handle.net/1843/42679
Full metadata record
DC FieldValueLanguage
dc.creatorSabrina Donato da Silvapt_BR
dc.creatorAndré de Souza Limapt_BR
dc.creatorThaís m. Mendespt_BR
dc.creatorLuciana Batista Nogueirapt_BR
dc.creatorTarcisio Passos Ribeiro de Campospt_BR
dc.date.accessioned2022-06-24T21:55:05Z-
dc.date.available2022-06-24T21:55:05Z-
dc.date.issued2019-
dc.citation.volume7pt_BR
dc.citation.spage01pt_BR
dc.citation.epage11pt_BR
dc.identifier.doi10.15392/bjrs.v7i2A.656pt_BR
dc.identifier.issn2319-0612pt_BR
dc.identifier.urihttp://hdl.handle.net/1843/42679-
dc.description.abstractNa verdade, o câncer ganhou uma dimensão maior e se tornou um problema de saúde pública mundial. A radioterapia (RT) é um tratamento de neoplasia e os aceleradores lineares de RT devem passar por um rigoroso controle de qualidade de dose. Phantoms de água ou água sólida podem ser usados ​​com esta intuição. Nos últimos anos, filmes radiocrômicos com composição tecidual equivalente têm sido amplamente utilizados como dosímetros na área médica. Neste trabalho a proposta foi analisar duas respostas distintas de filmes radiocrômicos em phantoms de água e água sólida, um espectro LINAC de feixe de 4MV. Os filmes solid water phantom, water phantom e radiocrômico EBT2 foram fixados em dois processos distintos de calibração. Os filmes foram expostos a um conjunto de doses absorvidas estabelecidas por unidades monitoras distintas (MU) especificadas no RT-center. Foram estabelecidas relações matemáticas entre o grau de intensidade de vermelho dos filmes digitalizados e a dose absorvida para ambos os métodos. Os coeficientes da função polinomial da curva de calibração foram determinados a partir do software Origin. A incerteza de ambos os processos foi analisada. A eficiência dos dois processos de calibração foi configurada. O ajuste da curva de calibração forneceu os coeficientes da equação de segunda ordem que relaciona a dose absorvida com a densidade óptica do filme. A incerteza quanto à calibração realizada em água e água sólida e a precisão do erro de dose estão de acordo com a literatura. Tanto a água quanto a água sólida foram eficazes na calibração e podem ser usadas em rotinas de medições de controle de qualidade. Os resultados mostram que os filmes radiocrômicos de EBT2 são adequados para calibração de dose em RT.pt_BR
dc.description.resumoActually, cancer has gained a larger dimension and become a global public health problem. Radiotherapy (RT) is a neoplasia treatment and RT linear accelerators must undergo a strict dose quality control. Water or solid water phantoms can be used with this intuit. In recent years, radiochromic films with equivalent tissue composition have been widely used as dosimeters in the medical field. In this work the proposal was to analyze two distinct radiochromic film responses in water and solid water phantoms, a LINAC spectrum of 4MV beam. Solid water phantom, water phantom and EBT2 Radiochromic films were set in two distinct process of calibration. Films were exposed to a set of absorbed doses established by distinct monitor units (MU) specified in RT-center. Mathematical relations between the degree of red-intensity from digitized films and the absorbed dose for both methods were established. The coefficients of the polynomial function of the calibration curve were determined from the Origin software. The uncertainty of both processes was analyzed. The efficiency of the two calibration processes was set up. The adjustment of the calibration curve provided the coefficients of the second-order equation that relates the dose absorbed with the optical density of the film. The uncertainty regarding the calibration performed in water and solid water and the dose-error accuracy are in agreement with the literature. Both water and solid water were effective in calibration and can be used in routines of quality-control measurements. The results show that EBT2-radiochromic films are suitable to for dose-calibration in RT.pt_BR
dc.format.mimetypepdfpt_BR
dc.languageengpt_BR
dc.publisherUniversidade Federal de Minas Geraispt_BR
dc.publisher.countryBrasilpt_BR
dc.publisher.departmentMED - DEPARTAMENTO DE ANATOMIA E IMAGEMpt_BR
dc.publisher.initialsUFMGpt_BR
dc.relation.ispartofBrazilian journal of radiation sciencespt_BR
dc.rightsAcesso Abertopt_BR
dc.subject.otherMedicinapt_BR
dc.subject.otherRadioterapiapt_BR
dc.titleComparison of the radiochromic EBT2 responses for4 MV LINAC in calibration processespt_BR
dc.title.alternativeComparação das respostas radiocrômicas de EBT2 para LINAC de 4 MV em processos de calibraçãopt_BR
dc.typeArtigo de Periódicopt_BR
dc.url.externawww.bjrs.org.br/revista/index.php/REVISTA/article/view/656pt_BR
Appears in Collections:Artigo de Periódico

Files in This Item:
File Description SizeFormat 
2019_Comparison of the radiochromic ebt2 responses for4 mv linac in calibration processes.pdf651.18 kBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.