Use este identificador para citar o ir al link de este elemento:
http://hdl.handle.net/1843/41267
Tipo: | Artigo de Periódico |
Título: | Activity monitor placed at the nonparetic ankle is accurate in measuring step counts during community walking in poststroke Individuals: a validation study |
Autor(es): | Noémie Duclos Larissa Tavares Aguiar Rachid Aissaoui Christina Danielli Coelho de Morais Faria Sylvie Nadeau Cyril Duclos |
Resumen: | Background: Different environmental factors may affect the accuracy of step-count activity monitors (AM). However, the validation conditions for AM accuracy largely differ from ecological environments. Objectives: To assess and compare the accuracy of AM in counting steps among poststroke individuals: during different locomotor tasks, with AM placed at the nonparetic ankle or hip, and when walking in a laboratory or inside a mall. Design: Validation study. Settings: Laboratory and community settings. Participants: Twenty persons with chronic hemiparesis, independent walkers. Methods: First session: participants performed level walking (6-minute walk test [6MWT]), ramps, and stairs in the laboratory with AM placed at the nonparetic ankle and hip. Second session: participants walked a mall circuit, including the three tasks, with AM placed at the nonparetic ankle. The sessions were video recorded. Main outcome measurements: Absolute difference between the steps counted by AM and the steps viewed on the video recordings (errors, %); occurrence of errors greater than 10%. Results: Median errors were similar for the 6MWT (0.86 [0.22, 7.70]%), ramps (2.17 [0.89, 9.61]%), and stairs (8.33 [2.65, 19.22]%) with AM at the ankle. Step-count error was lower when AM was placed at the ankle (8.33 [2.65, 19.22]%) than at the hip (9.26 [3.25, 42.63]%, P = .03). The greatest errors were observed among the slowest participants (≤0.4 m/s) on ramps and stairs, whereas some faster participants (>1 m/s) experienced the greatest error during the 6MWT. Median error was slightly increased in the mall circuit (2.67 [0.61, 12.54]%) compared with the 6MWT (0.50 [0.24, 6.79]%, P = .04), with more participants showing errors >10% during the circuit (7 vs 2, P = .05). Conclusions: Step counts are accurately measured with AM placed at the nonparetic ankle in laboratory and community settings. Accuracy can be altered by stairs and ramps among the slowest walkers and by prolonged walking tasks among faster walkers. Level of evidence: III. |
Asunto: | Caminhada - Medição Acidente vascular cerebral Avaliação |
Idioma: | eng |
País: | Brasil |
Editor: | Universidade Federal de Minas Gerais |
Sigla da Institución: | UFMG |
Departamento: | EEF - DEPARTAMENTO DE FISIOTERAPIA |
Tipo de acceso: | Acesso Aberto |
Identificador DOI: | https://doi.org/10.1002/pmrj.12080 |
URI: | http://hdl.handle.net/1843/41267 |
Fecha del documento: | 2019 |
metadata.dc.url.externa: | https://onlinelibrary.wiley.com/doi/10.1002/pmrj.12080 |
metadata.dc.relation.ispartof: | PM&R |
Aparece en las colecciones: | Artigo de Periódico |
archivos asociados a este elemento:
archivo | Descripción | Tamaño | Formato | |
---|---|---|---|---|
Activity monitor placed at the nonparetic ankle is accurate in measuring step counts during community walking.pdf | 468.46 kB | Adobe PDF | Visualizar/Abrir |
Los elementos en el repositorio están protegidos por copyright, con todos los derechos reservados, salvo cuando es indicado lo contrario.