Title: Avaliação dos serviços de medicina nuclear do estado de Minas Gerais: programa de garantia da qualidade dos equipamentos
Other Titles: Evaluation of the Nuclear Medicine facilities in Minas Gerais state: quality control program of equipments
Authors: Tadeu Takao Almodovar Kubo
Faber Henrique Zacarias Xavier
Rodrigo Modesto Gadelha Gontijo
Marcelo Mamede
Rodrigo Biancardi
Adriana Márcia Guimarães Rocha
Denia Romão Ferreira
Franciele Aquiles Anjos Silva
Jonathan Buenos Aires Assunção
Ederson Henrique Alves
Ana Flávia Batista Almeida
Nathália Fernandes Alves
Abstract: Resumo: Com a reformulação da norma CNEN-NN-3.05 em dezembro de 2013, os serviços de Medicina Nuclear (SMN) do Brasil têm que realizar um maior número de controles de qualidade para os equipamentos SPECT e PET. Entretanto, pouco se conhece sobre a realidade dos programas de garantia da qualidade desses serviços quanto a aplicação da nova norma. Assim, nesse contexto, o objetivo do presente trabalho foi avaliar o programa de controle da qualidade dos SMNs do estado de Minas Gerais. Todos os SMNs do estado de Minas Gerais foram convidados a participar do projeto. Desses, 34,48% (20 instalações) concordaram em participar do projeto, 50,00% (29 instalações) não responderam ao convite e 15,52% (9 instalações) declinaram da participação. Dessa forma, a partir de novembro de 2015, 20 equipamentos de SPECT e 2 PET/CT foram avaliados quanto à realização dos testes de controle de qualidade preconizados pela nova norma da CNEN. Os simuladores necessários à avaliação foram provenientes do Laboratório de Dosimetria e Controle de Qualidade da UFMG. Mesmo com os prazos estipulados pela CNEN quanto à implementação do programa de controle da qualidade nos SMN, mais de 50% dos serviços avaliados não implementaram os controles de qualidade, sendo a ausência de simuladores específicos o principal motivo da falha. Dentre os problemas encontrados nas instalações, os mais críticos foram: colimadores sem condições de utilização na rotina clínica, problemas de linearidade da imagem avaliada e valores de uniformidade da imagem superiores aos limites de aceitação. Problemas na uniformidade e na linearidade da imagem encontrados impactaram diretamente no desempenho de outros testes, como resolução espacial, desempenho SPECT, entre outros. De forma geral, os SMNs do estado de Minas Gerais, avaliados no presente estudo, encontram-se em condições de exequibilidade clínica.
Abstract: With the reformulation of the CNEN-NN-3.05 standard in December 2013, Brazil's Nuclear Medicine (NMS) services have to perform a greater number of quality controls for SPECT and PET equipment. However, little is known about the reality of the quality control programs of these services regarding the application of the new standard. Thus, in this context, the objective of the present study was to evaluate the quality control program of MNSs in the state of Minas Gerais. All NMSs in the state of Minas Gerais were invited to participate in the project. Of these, 34.48% (20 facilities) agreed to participate in the project, 50.00% (29 facilities) did not respond to the invitation and 15.52%(9 facilities) declined their participation. Thus, as of November 2015, 20 SPECT and 2 PET/CT equipments were evaluated for the performance of the quality control tests recommended by the new CNEN standard. The phantoms required for the evaluation came from the Laboratory of Dosimetry and Quality Control of UFMG. Even with the deadlines set by CNEN for the implementation of the quality control program in the NMSs, more than 50% of the evaluated services did not implement the quality controls, and the absence of specific phantoms is the main reason for the failure. Among the problems found in the installations, the most critical were: collimators with no conditions of use in the clinical routine, linearity problems of the evaluated image and values of image uniformity superior to the limits of acceptance. Problems in the uniformity and linearity of the image found directly impacted the performance of other tests, such as spatial resolution, SPECT performance, among others. In a general way, the NMSs in the state of Minas Gerais evaluated with the present study are in clinical feasible conditions.
Subject: Controle de qualidade
Medicina nuclear
Proteção radiológica
language: por
metadata.dc.publisher.country: Brasil
Publisher: Universidade Federal de Minas Gerais
Publisher Initials: UFMG
metadata.dc.publisher.department: MED - DEPARTAMENTO DE ANATOMIA E IMAGEM
Rights: Acesso Aberto
metadata.dc.identifier.doi: 10.29384/rbfm.2017.v11.n1.p35-39
URI: http://hdl.handle.net/1843/57380
Issue Date: 2017
metadata.dc.url.externa: https://www.rbfm.org.br/rbfm/article/view/404
metadata.dc.relation.ispartof: Revista Brasileira De Física Médica
Appears in Collections:Artigo de Periódico



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.