Título: Influence of detectors efficiency normalization on small animal PET image quality
Autor(es): Mateus Gesulado Carneirode Santana
Guilherme Cavalcante de Albuquerque Souza
Rodrigo Modesto Gadelha Gontijo
Bruno Melo Mendes
Andréa Vidal Ferreira
Resumen: A small animal positron emission tomograph (LabPET GE 4) image acquisition is done by 1536 independents channels. Differences in the rate counting of each channel must be corrected so as not compromise image quality. Equipment user manual recommends that normalization of the detectors efficiency be made as often as possible and always whenever there are hardware or software parameters changes - included, but not limited to channels parameters adjustments, electronic cards replacements, channels activation or inactivation or software update. This work evaluates the normalization effects on the image quality parameters. PET image acquisition and analysis were performed using recommended parameters by the NEMA NU 4-2008 standards. Image was reconstructed in different ways using different normalization files. The tests performed indicated that the image quality parameters do not vary significantly with different normalization data. Therefore, a high frequency routine of normalizations is not justified, suggesting a periodical frequency of one month for this procedure.
Asunto: Tomografia por Emissão de Pósitrons
Idioma: eng
País: Brasil
Editor: Universidade Federal de Minas Gerais
Sigla da Institución: UFMG
Departamento: MED - DEPARTAMENTO DE ANATOMIA E IMAGEM
Tipo de acceso: Acesso Aberto
Identificador DOI: 10.15392/bjrs.v7i3.861
URI: http://hdl.handle.net/1843/57382
Fecha del documento: 2019
metadata.dc.url.externa: https://www.bjrs.org.br/revista/index.php/REVISTA/article/view/861
metadata.dc.relation.ispartof: Brazilian Journal of Radiation Sciences
Aparece en las colecciones:Artigo de Periódico

archivos asociados a este elemento:
archivo Descripción TamañoFormato 
Influence of detectors efficiency normalization on pdfa.pdf808.78 kBAdobe PDFVisualizar/Abrir


Los elementos en el repositorio están protegidos por copyright, con todos los derechos reservados, salvo cuando es indicado lo contrario.