Título: Avaliação da vida restante em fadiga de instrumentos rotatórios de NiTi ProTaper após uso clínico múltiplo
Autor(es): Evandro Pires Vieira
primer Tutor: Maria Guiomar de Azevedo Bahia
primer Co-tutor: Vicente Tadeu Lopes Buono
primer miembro del tribunal : Antonio Paulino Ribeiro Sobrinho
Segundo miembro del tribunal: EUDES GONDIM JUNIOR
Resumen: A proposta desse estudo foi avaliar a vida restante em fadiga, bem como comparar os danos superficiais devido à fadiga estrutural, de instrumentos rotatórios de NiTi ProTaper utilizados na clínica para a limpeza e formatação do sistema de canais radiculares de cinco e oito molares. Quarenta e dois jogos de instrumentos ProTaper, tipo S1, S2, F1 e F2, foram analisados. Eles foram divididos em Grupo Controle (GC) com 12 jogos de instrumentos novos e três grupos experimentais: G1 e G2, cada um com 10 jogos de instrumentos, cada jogo usado clinicamente por um endodontista experiente com o sistema em cinco e oito molares, respectivamente; e G3, com 10 jogos de instrumentos, cada jogo usado clinicamente em 5 molares por estudantes de graduação inexperientes com o sistema. Antes e após o uso clínico, todos os instrumentos foram analisados pela presença de danos superficiais por microscopia ótica e três jogos de cada grupo de instrumentos usados, selecionados ao acaso, foram examinados por microscopia eletrônica de varredura (MEV). Os instrumentos foram em seguida testados em dispositivo de bancada para avaliação de sua resistência à fadiga. Os resultados obtidos foram analisados estatisticamente pelo teste ANOVA, com coeficiente de confiabilidade de 95%. Uma grande quantidade de microtrincas transversais foi observada em todos os instrumentos, bem como trincas longitudinais em menor número, indicando que eles sofreram fadiga estrutural. Foi observada uma redução da resistência à fadiga após o uso clínico para todos os instrumentos avaliados. A experiência do operador afetou a ocorrência deformação plástica e fratura das limas durante o uso clínico. Os resultados obtidos indicaram que o uso clínico múltiplo consome a vida em fadiga dos instrumentos, mas a vida restante em fadiga foi essencialmente a mesma nos instrumentos usados em cinco e oito molares.
Abstract: The purpose of this study was to evaluate the remaining fatigue life, as well as to compare surface damage due to structural fatigue, of rotary NiTi ProTaper instruments clinically used for cleaning and shaping the root canal system of five and eight molars. Forty-two sets of ProTaper instruments, type S1, S2, F1, and F2, were analyzed. They were divided into a Control Group (CG) with twelve sets of new instruments and three experimental groups: G1 and G2, each with ten sets of instruments, each set clinically used by an endodontist, with experience using the ProTaper system, in five and eight molars, respectively; and G3, with ten sets of instruments, each set clinically used in five molars by undergraduate students inexperienced with the system. Before and after the clinical use, all instruments were analyzed by optical microscopy for the presence of surface damage and three sets of each group of used instruments, randomly selected, were examined by scanning electron microscopy (SEM). The instruments were then tested in a bench device for evaluation of their fatigue resistance. The results obtained were statistically analyzed by one way ANOVA tests, at the 95% confidence level. A large amount of transverse microcracks were observed in all instruments, as well as a smaller number of longitudinal microcracks, indicating that they suffered structural fatigue. After the clinical use, a decrease in fatigue resistance was observed for all instruments evaluated. Operator experience affected the occurrence of plastic deformation and fracture of the files during shaping. The results obtained indicate that multiple clinical uses consume the fatigue life of the instruments, but the remaining fatigue life was essentially the same for instruments used in five and eight molars.
Asunto: Canal radicular Tratamento
Instrumentos e aparelhos odontológicos
Idioma: Português
Editor: Universidade Federal de Minas Gerais
Sigla da Institución: UFMG
Tipo de acceso: Acesso Aberto
URI: http://hdl.handle.net/1843/ZMRO-77FLW3
Fecha del documento: 23-feb-2007
Aparece en las colecciones:Dissertações de Mestrado

archivos asociados a este elemento:
archivo Descripción TamañoFormato 
evandro_folhas_iniciais.corrigida_final.pdf7.33 MBAdobe PDFVisualizar/Abrir


Los elementos en el repositorio están protegidos por copyright, con todos los derechos reservados, salvo cuando es indicado lo contrario.