Computational analysis of 'Dose/Collision Kerma' relationship and lateral boundary in Stereotatic circular fields using EGSnrc.

dc.creatorReginaldo Gonçalves Leão Junior
dc.creatorRômulo Verdolin de Sousa
dc.creatorArno Heeren de Oliveira
dc.creatorHugo Lemos Leonardo Silva
dc.creatorArnaldo Prata Mourão
dc.date.accessioned2022-10-18T14:59:01Z
dc.date.accessioned2025-09-09T01:29:06Z
dc.date.available2022-10-18T14:59:01Z
dc.date.issued2016
dc.description.abstractEste trabalho teve como objetivo principal a obtenção de dados de pequenos campos de raios X que evidenciassem as hipóteses apontadas como dificultadoras para a dosimetria destes. Foi realizadaa verificação da compatibilidade entre o limitedo campo dosimétrico com o campo geométrico e a verificação da relação kerma-dose segundo oesperado paracampos convencionais. Materiais e Métodos: Simulações computacionais de campos menores que 5x5 cm² foram realizadasutilizando o método de Monte Carlopor meioda aplicação egs_chamber, esta derivada do código de transporte de radiação EGSnrc. Como fontesde partículas,foram utilizados phase space filesde um modelo de cabeçote do acelerador linear CLINAC 2100acoplado a cones deRadiocirurgia Estereotáxica. Resultados: As simulações sugeriram a existência de um patamar, próximo de 8%,nas discrepâncias entre o FWHMda Dosee o diâmetro nominal do campo, indicando um decréscimo destes valores para campos de diâmetrosmenoresque 12 mm e maiores que 36mm.Simultaneamente,asdiscrepâncias dose-kerma em profundidade atingiram valoressuperiores a14% no caso onde o fenômeno foi mais significativo. Conclusão: Os dadosda relação kerma-dosemostraram que,de fato,o comportamento dos pequenos campos destoa daquele esperado para campos convencionais, e que as convenções dosimétricas tradicionais não se aplicam a tais campos,sendo necessária uma abordagem especializada para as técnicas que os empregam.Além disto, a existência do já referido patamar de discrepâncias, juntamente com a diminuição destas em campos de diâmetro menor que 15 mm, constituem um achado notável.
dc.format.mimetypepdf
dc.identifier.doihttps://doi.org/10.29384/rbfm.2016.v10.n1.p2-8
dc.identifier.issn1984-9001
dc.identifier.urihttps://hdl.handle.net/1843/46315
dc.languageeng
dc.publisherUniversidade Federal de Minas Gerais
dc.relation.ispartofRevista Brasileira de Física Médica
dc.rightsAcesso Aberto
dc.subjectEngenharia nuclear
dc.subject.otherSnall fields
dc.subject.otherDosimetry
dc.subject.otherMonte Carlo
dc.subject.otherEGSnrc
dc.subject.otherPequenos Campos
dc.subject.otherDosimetria
dc.titleComputational analysis of 'Dose/Collision Kerma' relationship and lateral boundary in Stereotatic circular fields using EGSnrc.
dc.typeArtigo de periódico
local.citation.epage8
local.citation.issue1
local.citation.spage2
local.citation.volume10
local.description.resumoThis work aimed to obtain data from small fields of X-rays that evidence of the hypotheses cited as cause of difficulties for the dosimetry of these. For this purpose, the verification of compatibility between the dosimetric boundary offield and the geometricsize offield, was performed. Also,theverification of kerma dose according to the expected relationship for conventional fieldswas made.Materials and Methods: Computer simulations of smaller fields 5x5 cm² were performed,using the Monte Carlo method by egs_chamber application, this derived from EGSnrc radiation transport code.As particulate sources were used phase spacefilesof a Clinac 2100head modelcoupled to cones Stereotactic Radiosurgery.Results: The simulations suggested the existence of a plateau in discrepancies between thedoseFWHM and the nominal diameter of the fieldclose to 8%. These simulations also indicated a decrease of these values for fields with diameterssmallerthan 12mm and larger than 36mm.Simultaneously, the dose kerma differences in depth reached values higher than 14% in the case where the phenomenon is more significant.Conclusion: The data showed that in fact the behavior of small fields clashes with that expected for conventional fields, and that the traditional dosimetric conventions do not apply to such fields requiring a specialized approach to the techniques that employ them.Furthermore, the existence of the aforementioned plateauofdiscrepancies, along with the decrease thereof in less than 15 mm diameter fields constitute a remarkable finding.
local.publisher.countryBrasil
local.publisher.departmentENG - DEPARTAMENTO DE ENGENHARIA NUCLEAR
local.publisher.initialsUFMG
local.url.externahttps://www.rbfm.org.br/rbfm/article/view/355

Arquivos

Pacote original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
Computational analysis of 'DoseCollision Kerma' relationship and - PDFA.pdf
Tamanho:
499.43 KB
Formato:
Adobe Portable Document Format

Licença do pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
License.txt
Tamanho:
1.99 KB
Formato:
Plain Text
Descrição: