Constancy tests and quality assurance of the activimeters used in a radiopharmaceutical production unit

dc.creatorLeonardo Tafas Constantino Nascimento
dc.creatorFlávia Mesquita Costa
dc.creatorJuliana Batista da Silva
dc.creatorRodrigo Modesto Gadelha Gontijo
dc.creatorAndréa Vidal Ferreira
dc.creatorMarcelo Henrique Mamede Lewer
dc.date.accessioned2023-08-01T21:32:57Z
dc.date.accessioned2025-09-08T22:57:52Z
dc.date.available2023-08-01T21:32:57Z
dc.date.issued2019
dc.format.mimetypepdf
dc.identifier.doi10.15392/bjrs.v7i2A.588
dc.identifier.issn23190612
dc.identifier.urihttps://hdl.handle.net/1843/57336
dc.languageeng
dc.publisherUniversidade Federal de Minas Gerais
dc.relation.ispartofInternational Nuclear Atlantic Conference
dc.rightsAcesso Aberto
dc.subjectRadioisótopos
dc.subject.otherActivimeters
dc.subject.otherRadionuclides
dc.titleConstancy tests and quality assurance of the activimeters used in a radiopharmaceutical production unit
dc.typeArtigo de evento
local.citation.epage13
local.citation.issue2A
local.citation.spage1
local.description.resumoActivimeters (or dose calibrators) are essential instruments to verify activity of radiopharmaceutical after production and also before the dose administration in humans or animals for molecular imaging. The efficiency and safety measurements depend on, beside other factors, constancy tests and quality assurance. Thereby, the aim of this work was to perform constancy tests and quality assurance in the activimeters of the UPPR/CDTN,based on the CNEN-NN 3.05 Brazilian standard and the manufacturer's manual. Physical inspection, auto zero,background check, camera voltage, data check and constancy test were done. In addition, accuracy and precision tests were performed using a set of standard certified radioactive sources (57Co, 133Ba and 137Cs), according to the CNEN NN 3.05 Brazilian standard. Linearity test was also performed to evaluate the response of the equipment in over the entire range of activities used in routine. The equipments are periodically submitted to the quality control tests and the results were compared. After performing the proposed tests it is possible to conclude that activimeters are in accordance with the requirements of the CNEN standard and manufacturer´s manual. A quality control checklist was prepared to guide users and to record the results of quality assurance testing to monitor the equipment performance. This initiative is part of the quality assurance program implemented at UPPR.
local.identifier.orcidhttps://orcid.org/0000-0001-6479-3087
local.publisher.countryBrasil
local.publisher.departmentMED - DEPARTAMENTO DE ANATOMIA E IMAGEM
local.publisher.initialsUFMG

Arquivos

Pacote original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
CONSTANCY TESTS AND QUALITY ASSURANCE OF THE pdfa.pdf
Tamanho:
377.12 KB
Formato:
Adobe Portable Document Format

Licença do pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
License.txt
Tamanho:
1.99 KB
Formato:
Plain Text
Descrição: