Mean glandular dose to patients from stereotactic breast biopsy procedures
Carregando...
Data
Título da Revista
ISSN da Revista
Título de Volume
Editor
Universidade Federal de Minas Gerais
Descrição
Tipo
Artigo de periódico
Título alternativo
Primeiro orientador
Membros da banca
Resumo
The aim of this work is to study the radiation doses delivered to a group of patients that underwent a stereotactic breast biopsy (SBB) procedure. Mean glandular doses (MGD) were estimated from the air-kerma measured at the breast surface entrance multiplying by specific conversion coefficients (DgN) that were estimated using Monte Carlo simulations. DgN were calculated for the 0° and ±15°
projections used in SBB and for the particular beam quality. Data on 61 patients were collected showing that a typical SBB procedure is composed by ten images. MGD was on average (4 ± 2) mGy with (0.38 ± 0.06) mGy per image. The use of specific conversion coefficients instead of typical DgN for mammography/tomosynthesis yields to obtain MGD values for SBB that are around a 65% lower
on average.