Concurrent validity of handgrip strength between the jamar and bulb dynamometers in women with rheumatoid arthitis

dc.creatorLygia Paccinilustosa
dc.creatorKarla Gonçalves Diogo
dc.creatorGiane Amorim Ribeiro-samora
dc.creatorAdriana Maria Kakehasi
dc.creatorMariana Asmar Alencar
dc.date.accessioned2023-12-18T23:24:38Z
dc.date.accessioned2025-09-09T00:14:58Z
dc.date.available2023-12-18T23:24:38Z
dc.date.issued2020
dc.description.abstractIntrodução: Artrite reumatóide (AR) é uma doença progressiva que causa deformação e comprometimento articular. A força de preensão palmar (FPP) tem sido usada por vários profissionais de saúde, na prática clínica, para avaliar força muscular e desempenho. Objetivo: Analisar a concordância das medidas de força de preensão palmar usando o dinamômetro pneumático (Bulbo) e hidráulico (Jamar) em mulheres com artrite reumatoide. Método: As medidas de FPP pelos dois dinamômetros seguiram as normas da American Society of Hand Therapists. A concordância entre as medidas foi por meio do método de Bland-Altman. Nível de significância de 5%. Resultados: Participaram 41 mulheres (60,63 ± 8,35 anos). A análise entre os dois dinamômetros demonstrou não haver concordância (bias = 9,04, p < 0,01), com uma relação linear entre a diferença das medidas e a média (r = 0,73, p < 0,01) e, os limites de concordância foram muito extensos (-2,74 to 20,81). Conclusão: Como a artrite reumatoide compromete as mãos com consequente diminuição da FPP, é importante explorar melhor este tema para definir a melhor medida para a prática clínica nos diferentes níveis de atenção à saúde. No entanto, devido ao grande número de tipo de dinamômetros, sugere-se novos estudos sobre a concordância destas medidas em populações de diferentes condições.
dc.format.mimetypepdf
dc.identifier.doi10.1590/1980-5918.033.ao19
dc.identifier.issn19805918
dc.identifier.urihttps://hdl.handle.net/1843/62068
dc.languageeng
dc.publisherUniversidade Federal de Minas Gerais
dc.relation.ispartofFisioterapia em Movimento
dc.rightsAcesso Aberto
dc.subjectArthritis, Rheumatoid
dc.subjectArthritis
dc.subjectMuscle Strength
dc.subjectMuscle Strength Dynamometer
dc.subject.otherArthritis, Rheumatoid
dc.subject.otherArthritis
dc.subject.otherMuscle Strength
dc.subject.otherMuscle Strength Dynamometer
dc.titleConcurrent validity of handgrip strength between the jamar and bulb dynamometers in women with rheumatoid arthitis
dc.typeArtigo de periódico
local.citation.epage7
local.citation.spage1
local.citation.volume33
local.description.resumoIntroduction: Rheumatoid arthritis (RA) is a progressive disease that causes deformation and constant joint damage. Handgrip strength (HGS) has been used by several health professionals in clinical practice as a mechanism for assessing muscle strength and overall performance. Objective: To perform the concordance analysis of handgrip strength measurements using a pneumatic dynamometer (Bulb) and a hydraulic dynamometer (Jamar) in women with rheumatoid arthritis. Method: The HGS measurements by the two dynamometers followed the norms of the American Society of Hand Therapists. The concordance between measurements was performed by the Bland-Altman method. Significance level was 5%. Results: In total,41 women (60.63 ± 8.35 years) participated in the study. Analysis showed that the measures between the two dynamometers were not concordant (bias = 9.04, p < 0.01), there was a linear relationship between the difference of the measures and the mean (r = 0.73, p < 0.01), and the limits of agreement were very extensive (−2.74 to 20.81). Conclusion: Assuming that the rheumatoid arthritis compromises the hands, with a consequent decrease in HGS, further exploration of the subject is suggested in future studies to define the best measure for clinical practice at the different levels of health care. However, since there are many different dynamometers, we suggest to better explore the agreement between the measurements obtained by them in populations under different conditions.
local.publisher.countryBrasil
local.publisher.departmentMED - DEPARTAMENTO DE APARELHO LOCOMOTOR
local.publisher.departmentMED - DEPARTAMENTO DE CLÍNICA MÉDICA
local.publisher.initialsUFMG
local.url.externahttps://doi.org/10.1590/1980-5918.033.AO19

Arquivos

Pacote original

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
Concurrent validity of handgrip strength between the jamar pdfa.pdf
Tamanho:
270.89 KB
Formato:
Adobe Portable Document Format

Licença do pacote

Agora exibindo 1 - 1 de 1
Carregando...
Imagem de Miniatura
Nome:
License.txt
Tamanho:
1.99 KB
Formato:
Plain Text
Descrição: