Please use this identifier to cite or link to this item: http://hdl.handle.net/1843/35962
Type: Dissertação
Title: Desenvolvimento de modelo para simulação de sistema de aspiração endotraqueal em modo contínuo ou intermitente
Other Titles: Development of a model for simulation of endotracheal suctioning system in continuous or intermittent mode
Authors: Fabiano Viana Oliveira da Cunha Médice
First Advisor: Leonardo Antônio Borges Tôrres
First Co-advisor: Alcendino Cândido Jardim Neto
First Referee: Henrique Resende Martins
Second Referee: Gabriel Casulari da Motta Ribeiro
Abstract: A aspiração endotraqueal é um recurso clínico comumente utilizado para a higienização brônquica de pacientes sob ventilação artificial que pode acarretar diversas complicações e até o óbito. Com o intuito de minimizar danos aos pacientes simulados pela queda de pressão durante esse processo, esse trabalho visa propor um novo procedimento, onde se utiliza a aspiração intermitente. Com a aspiração intermitente, espera-se que um menor volume de ar seja retirado do paciente simulado, acarretando assim uma menor queda de pressão. Tendo em vista a importância de se avaliar o impacto da interação ventilador-aspirador no sistema respiratório, um dos objetivos deste trabalho é a validação de um modelo computacional de um sistema fechado de aspiração traqueal. Para tanto, compararam-se as pressões obtidas pelo modelo computacional àquelas resultantes da aspiração em um modelo físico de pulmão, empregando diferentes combinações de sonda, tubo orotraqueal, pressões ventilatórias e de aspiração. A diferença entre as pressões alveolares obtidas com os modelos computacional e físico não foram superiores a 0,3 cmH2O; havendo uma correlação entre os sinais superior a 0,999. Portanto, considerou-se o modelo computacional adequado à representação mecânica alveolar na condição proposta. Com o modelo digital validado, implementou-se uma aspiração com parâmetros de 4 tipos de pacientes simulados, sendo um saudável, um com doença restritiva, um com doença obstrutiva e um crítico com ambas doenças; e comparou-se a simulação com aspiração contínua com a aspiração intermitente. Utilizando a aspiração intermitente, houve uma melhora dos pacientes simulados com doenças respiratórias onde se obtiveram uma pressão média alveolar maior e uma diminuição de queda máxima de pressão durante a aspiração nos pacientes.
Abstract: Endotracheal Suctioning is a clinical resource commonly used for the bronchial hygiene of patients under artificial ventilation that can lead to several complications and even death. Seeking to minimize damage to the simulated patient during this process, this work proposes a new procedure, where intermittent suctioning is used. When intermittent suctioning is used, it is expected that a smaller volume of air will be removed from the simulated patient, resulting in a lower pressure drop. Considering the importance of assessing the impact of the Mechanical Ventilation-Suctioning interaction on the respiratory system, one of the objectives of this work is the validation of a computational model of a closed tracheal suctioning system. For that, the pressures obtained by the computational model were compared to those resulting from suctioning in a physical lung model, using combinations of different sizes of catheters, orotracheal tubes and ventilatory and suctioning pressures. The difference between the alveolar pressures obtained with the computational and physical models was not higher than 0.3 cmH2O; with a correlation between the signals greater than 0.999. Therefore, the computational model was considered adequate for the alveolar mechanical representation in the proposed condition. With the validated digital model, a suctioning with parameters of 4 type of simulated patients was implemented (one healthy, one with restrictive disease, one with obstructive disease and one critical with both diseases) and the simulation with continuous suctioning was compared with intermittent suctioning. When comparing continuous and intermittent suctioning, there was an improvement in simulated patients with respiratory diseases where a higher mean alveolar pressure and a lower maximum pressure drop during suctioning in these patients were obtained.
Subject: Engenharia elétrica
Bioengenharia
Modelos computacionais
language: por
metadata.dc.publisher.country: Brasil
Publisher: Universidade Federal de Minas Gerais
Publisher Initials: UFMG
metadata.dc.publisher.department: ENG - DEPARTAMENTO DE ENGENHARIA ELÉTRICA
metadata.dc.publisher.program: Programa de Pós-Graduação em Engenharia Elétrica
Rights: Acesso Aberto
URI: http://hdl.handle.net/1843/35962
Issue Date: 26-Oct-2020
Appears in Collections:Dissertações de Mestrado

Files in This Item:
File Description SizeFormat 
Dissertação do Mestrado - rev Final - Versão Impressão - Versão Final.pdf4.03 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.