Title: Análise de elementos finitos da distribuição de tensões equivalentes nos pinos de schanz durante o uso do distrator femoral de fraturas
Other Titles: Finite element analysis of the equivalent stress distribution in Schanz screws during the use of a femoral fracture distractor
Authors: Vincenzo Giordano
Alexandre Leme Godoy'santos
William Dias Belangero
Robinson Esteves Santos Pires
Pedro José Labronici
Hilton Augusto Koch
Abstract: Avaliar o comportamento do estresse mecânico e da deformação elástica exercida natransição rosca-talo liso dos pinos de Schanz do distrator femoral de fraturas em montagens com diferentes distâncias de ancoragem dos pinos na cortical óssea de entrada através de estudo da distribuição e da localização de tensões no corpo de prova.Feita análise de elementos finitos 3 D para a avaliação da distribuição das tensões equivalentes em um pino de Schanz fixado de modo bicortical e ortogonal a um osso tubular,em dois padrões de montagem: (1) transição rosca-talo liso distante 20mmda ancoragemdos pinos de Schanz na cortical de entrada e (2) transição rosca-talo liso distante 3 mmda ancoragem dos pinos de Schanz na cortical de entrada. Foram feitas simulações e manteve- -se a mesma direção da carga e a mesma distância do vetor força em relação ao centro do osso hipotético. A carga aplicada, sua direção e a distância ao centro do osso foram constantes durante as simulações para manter o momento fletor igualmente constante. Os cálculos apresentados demonstraram comportamento linear durante todo o experimento.Verificou-se que o modelo com uma distância de 20mm entre a ancoragem dos pinos de Schanz na cortical óssea de entrada e a transição rosca-talo liso reduziu o risco de rupturaou fadiga do material durante a aplicação de cargas estáticas constantes. Nesse modelo,as forcas máximas observadas foram superiores (350 Mpa). A análise do comportamento do estresse mecânico e da deformação elástica exercida na transição rosca-talo liso dos pinos de Schanz do distrator femoral de fraturas mostrou que distâncias maiores entre a ancoragem dos pinos na cortical óssea de entrada e a transição rosca-talo liso dos pinos de Schanz. permitem menor estresse mecânico na transic¸ão rosca-talo liso e maior grau de deformação elástica do material e minimizam quebra ou fadiga. A distância sugerida é de 20mm
Abstract: To evaluate the mechanical stress and elastic deformation exercised in the thread/shaft transition of Schanz screws in assemblies with different screw anchorage distances in the entrance to the bone cortex, through the distribution and location of tension in the samples. An analysis of 3 D finite elements was performed to evaluate the distribution of the equivalent stress (triple stress state) in a Schanz screw fixed bicortically and orthogonally to a tubular bone, using two mounting patterns: (1) thread/shaft transition located 20mm from the anchorage of the Schanz screws in the entrance to the bone cortex and (2) thread/shaft transition located 3 mm from the anchorage of the Schanz screws in entrance to the bone cortex. The simulations were performed maintaining the same direction of loading and the same distance from the force vector in relation to the center of the hypothetical bone. The load applied, its direction, and the distance to the center of the bone were constant during the simulations in order to maintain the moment of flexion equally constant. The present calculations demonstrated linear behavior during the experiment. It was found that the model with a distance of 20mm between the Schanz screws anchorage in the entrance to the bone cortex and the thread/shaft transition reduces the risk of breakage or fatigue of the material during the application of constant static loads; in this model. the maximum forces observed were higher (350 Mpa). The distance between the Schanz screws anchorage at the entrance to the bone cortex and the smooth thread/shaft transition of the screws used in a femoral distractor during acute distraction of a fracture must be farther from the entrance to the bone cortex, allowing greater degree of elastic deformation of the material,lower mechanical stress in the thread/shaft transition, and minimized breakage or fatigue. The suggested distance is 20mm.
Subject: Análise de elementos finitos
Pinos ortopédicos
Fraturas do Fêmur
Fixadores Externos
Resultado do Tratamento
language: por
metadata.dc.publisher.country: Brasil
Publisher: Universidade Federal de Minas Gerais
Publisher Initials: UFMG
metadata.dc.publisher.department: MED - DEPARTAMENTO DE APARELHO LOCOMOTOR
Rights: Acesso Aberto
metadata.dc.identifier.doi: https://doi.org/10.1016/j.rbo.2016.06.009
URI: http://hdl.handle.net/1843/56975
Issue Date: 23-Jan-2017
metadata.dc.url.externa: https://www.sciencedirect.com/science/article/pii/S0102361616302752?via%3Dihub
metadata.dc.relation.ispartof: Revista Brasileira de Ortopedia
Appears in Collections:Artigo de Periódico



Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.