Please use this identifier to cite or link to this item: http://hdl.handle.net/1843/57386
Type: Artigo de Periódico
Title: Influence of image reconstruction protocols on the image quality of a small animal PET scanner using 18F and 11C
Authors: João Vitor do Carmo Barbosa
Rodrigo Modesto Gadelha Gontijo
Guilherme Cavalcante de Albuquerque Souza
Bruno Melo Mendes
Juliana Batista da Silva
Marcelo Mamede
Andréa Vidal Ferreira
Abstract: The small animal positron emission tomography (PET) scanner from Molecular Imaging Laboratory (LIM/CDTN) is dedicated to pre-clinical studies on new 18F and 11C-based radiopharmaceuticals and to the development of novel applications for traditional radiopharmaceuticals. Quality control tests recommended by the NEMA publication NU 4-2008 are routinely carried out to ensure the proper performance of the PET scanner. The aim of this work was to evaluate the influence of image reconstruction protocols on the image quality, accuracy of attenuation and scatter correction parameters for 18F and 11C PET images. 18F-FDG and 11C-PK1122 PET images of the image quality phantom were acquired and then reconstructed using different reconstruction protocols varying algorithms (FBP, MLEM-3D, OSEM-3D), resolution mode (high/standard) and number of iterations (10 to 150). Uniformity, spill-over ratio (SOR) and recovery coefficient (RC) tests were performed for each reconstructed image according to NEMA NU 4-2008. FBP based protocol generated noisier images compared to iterative algorithms (MLEM-3D or OSEM-3D) based protocols. The increase in the number of iterations resulted in higher standard deviation of the analyzed parameters for all images. MLEM-3D and OSEM-3D based protocols generates similar results when the number of iterations and resolution mode were identical. SOR and RC mean values remained stable when the number of iterations ranged from 40 to 150. This study allowed the evaluation of different image reconstruction protocols on important parameters of 18F and 11C PET image quality. Additionally, standard protocols to be adopted in LIM/CDTN for 18F and 11C images reconstruction in preclinical studies were defined.
Subject: Tomografia por Emissão de Pósitrons
Diagnóstico por Imagem
language: eng
metadata.dc.publisher.country: Brasil
Publisher: Universidade Federal de Minas Gerais
Publisher Initials: UFMG
metadata.dc.publisher.department: MED - DEPARTAMENTO DE ANATOMIA E IMAGEM
Rights: Acesso Aberto
metadata.dc.identifier.doi: 10.15392/bjrs.v9i3.1734
URI: http://hdl.handle.net/1843/57386
Issue Date: 20-Sep-2021
metadata.dc.url.externa: https://bjrs.org.br/revista/index.php/REVISTA/article/view/1734
metadata.dc.relation.ispartof: Brazilian Journal of Radiation Sciences
Appears in Collections:Artigo de Periódico

Files in This Item:
File Description SizeFormat 
Influence of image reconstruction protocols on the pdfa.pdf1.18 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.