GC-MS/MS versus LC-MS/MS for unambiguous identification of α and β-trenbolone in bovine urine
Carregando...
Data
Autor(es)
Título da Revista
ISSN da Revista
Título de Volume
Editor
Universidade Federal de Minas Gerais
Descrição
Tipo
Artigo de periódico
Título alternativo
Primeiro orientador
Membros da banca
Resumo
This work involves the optimization of LC-MS/MS and GC-MS/MS methods for determination of α and β-trenbolone in bovine urine. The parameters recovery, repeatability and intermediate precision in both methods met the acceptance criteria of the Codex Alimentarius: recovery between 83.5-114% and repeatability and intermediate precision with RSD between 5.1-14.5% and 9.8-25.5% respectively, for concentrations higher than 1 μg kg−1. These methods were applied to the analyses of 5 real samples suspected of contamination. The analyses showed that derivatization with MSTFA/I2 followed by pure MSTFA, was an important tool to prevent false positive results. The use of microwave assisted derivatization reduced significantly the analysis time by GC-MS/MS.
Abstract
Assunto
Urina, Análise, Cromatografia de gás, Espectrometria de massa, Cromatografia líquida, Bovino Saúde
Palavras-chave
Bovine urine, Gas chromatography tandem mass spectrometry, Liquid chromatography tandem mass spectrometry, Microwave assisted derivatization, Trenbolone