GC-MS/MS versus LC-MS/MS for unambiguous identification of α and β-trenbolone in bovine urine

Carregando...
Imagem de Miniatura

Data

Título da Revista

ISSN da Revista

Título de Volume

Editor

Universidade Federal de Minas Gerais

Descrição

Tipo

Artigo de periódico

Título alternativo

Primeiro orientador

Membros da banca

Resumo

This work involves the optimization of LC-MS/MS and GC-MS/MS methods for determination of α and β-trenbolone in bovine urine. The parameters recovery, repeatability and intermediate precision in both methods met the acceptance criteria of the Codex Alimentarius: recovery between 83.5-114% and repeatability and intermediate precision with RSD between 5.1-14.5% and 9.8-25.5% respectively, for concentrations higher than 1 μg kg−1. These methods were applied to the analyses of 5 real samples suspected of contamination. The analyses showed that derivatization with MSTFA/I2 followed by pure MSTFA, was an important tool to prevent false positive results. The use of microwave assisted derivatization reduced significantly the analysis time by GC-MS/MS.