Abstract: O presente estudo avaliou a influência do NaOCl e da ação de corte na resistência à fadiga de instrumentos de NiTi ProFile acionados a motor, submetidos à formatação de 10 canais radiculares curvos e posteriormente ensaiados em fadiga, utilizando um dispositivo de bancada para fixação do sistema de acionamento dos instrumentos eum canal artificial de aço carbono com raio de curvatura de 4mm e ângulo de curvatura de 45°. O número de ciclos até a fratura foi determinado a partir do tempo de ensaio e da velocidade de rotação dos instrumentos, 250 rpm. Oitenta instrumentos novos, 20/.06, 30/.04, 25/.04 e 20/.04, foram divididos aleatoriamente em 4 Gruposcom 20 amostras por grupo (cinco de cada calibre), nas seguintes condições: Grupo Controle constituído de instrumentos sem uso; Grupo 1 constituído de instrumentos imersos por 24 horas em solução de NaOCl a 5,25%; Grupo 2 constituído de instrumentos utilizados previamente na formatação de 10 canais radiculares curvos dedentes extraídos, empregando água deionizada como solução irrigante; e Grupo 3 constituídos de instrumentos empregados na formatação de 10 canais radiculares curvos utilizando NaOCl a 5,25% como solução irrigante. A imersão em NaOCl não alterou a superfície dos instrumentos, nem sua resistência à fadiga, entretanto a açãode corte resultante da formatação dos canais radiculares sob condições de deformação cíclica foi o fator crucial na redução do número de ciclos até a fratura. (Testes t de Student e Kruskal-Wallis p< 0,05). A resistência à fadiga diminuiu com o aumento do calibre dos instrumentos em todos os grupos. Os instrumentos do Grupo 3 foram estatisticamente mais rápidos na formatação dos canais em comparação com oGrupo 2 (Teste t de Student p< 0,05), havendo uma tendência de apresentarem um maior número de ciclos até a fratura que o Grupo 2, mas estes resultados não foram estatisticamente significativos (Testes t de Student e Kruskal-Wallis p> 0,05). O emprego do NaOCl durante a formatação dos canais levou a um menor tempo de instrumentação, o que pode constituir fator importante no aumento da vida útil dosinstrumentos.
Abstract: The present study evaluated the influence of sodium hypochlorite (NaOCl) and of the cutting action on the fatigue resistance of ProFile rotary NiTi endodontic instruments, employed for shaping 10 curved root canals and then tested in fatigue, using a stand for fixation of the hand piece and an carbon steel artificial canal with radius of 4mm andangle of curvature of 45º. The number of cycles to failure was determined in terms of the measured test time and the rotation speed, 250rpm. Eighty new instruments, 20/.06, 30/.04, 25/.04 and 20/.04, were randomly divided into a four Groups, with 20 samples in each Group, (five of each size), in the following conditions: control Group, with 5 non-used instruments of each type; Group 1 with the same amount ofinstruments, which were immersed for 24 hours in a solution of NaOCl 5.25%; Group 2, also with 5 instruments of each type, previously employed for preparing 10 curved canals of extracted teeth using deionized water as irrigant; and Group 3, with the same amount of instruments, after their use in formatting 10 curved canals using NaOCl5.25% as irrigant. The immersion in NaOCl did not affect the surface of the instruments nor their fatigue resistance, whereas the cutting action resulting of their use in formatting curved canals was the critical factor to reduce the number of cycles to failure (Students t test and Kruskal-Wallis p < 0,05)e. It has been observed that the fatigue resistance decreased with the increase in instrument size, in all Groups. The instruments of Group 3 were statistically faster in shaping the root canals than those inGroup 2 (Students t test p< 0,05). There was a tendency of the instruments of Group 3 to show greater fatigue resistance than those of Group 2, although these results were not statistically significant (Students t test and Kruskal-Wallis p > 0,05). The use of NaOCl during shaping of curved root canals reduced the time spent in the instrumentation and this can be an important factor to increase the useful life of theinstruments.
Subject: Instrumentos e aparelhos odontológicos
Ligas de níquel-titânio
Endodontia
language: Português
Publisher: Universidade Federal de Minas Gerais
Publisher Initials: UFMG
Rights: Acesso Aberto
URI: http://hdl.handle.net/1843/ZMRO-7JCP8Z
Issue Date: 27-Feb-2003
Appears in Collections:Dissertações de Mestrado

Files in This Item:
File Description SizeFormat 
disserta__o_mestrado_renata_martins.pdf9.38 MBAdobe PDFView/Open


Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.